Abstract:
Aim: The aim of this study is to evaluate long-term durability and effectiveness of the adjustable transobturator male system (ATOMS).
Materials and methods: The retrospective multicenter Iberian ATOMS study (n = 215) was updated to evaluate long-term continence status, complications, explants, and secondary treatments. Mean follow-up from surgery to March 2020 was 60.6 ± 18.4 months (range, 39-91). Eleven patients deceased of an unrelated causes. Kaplan-Meier curves were performed to evaluate device durability and incontinence free of recurrence interval. The multivariate analysis defined the population at risk of device explant.
Results: A total of 155 patients were dry at the last follow-up visit (72.1%); 99 (46%) used no pads and 56 (26%) used a security pad/day with urine loss less than 10 mL; 96% of dry patients after adjustment remained free of incontinence 1 year later, 93.6% 2 years later, 91.1% 3 years later, 89.2% 5 years later, and 86.7% 8 years later. Complications during follow-up occurred in 43 of 215 (20%). In total, 25 (11.6%) devices were explanted and causes were inefficacy 11 (44%), inefficacy and pain 3 (12%), port erosion 10 (40%), and wound infection 1 (4%). The secon...